User Manual

2-16 ADS TRITON+ Manual
Peak Combo Sensor
Ultrasonic Depth Sensor
To provide the depth of the flow, the sensor se nds an ultrasonic
signal upward that bounces off the water surface air interface.
The travel time fro m the s ensor to the flow surface and back is
converted to distance, which translates into the depth of the flow.
Upward ultrasonic depth signals bouncing off water surface
Pressure Depth Sensor
A pressure transducer in t he Peak Combo Sensor also measures the
depth of the flow over the sensor. The monitor can use this data for
comparison with other depth measurement devices in less tha n ful l-
pipe conditions. The pressure depth sensor also can measure dep ths
greater than full pipe that may extend up into t he manhole during
surcharge conditions.
Water pressure impacts a diaphragm in the t ransduc er that varies an
electrical current. This current is digitized by a sensor processor