User Manual
D-2 ADS TRITON+ Manual
intrinsically safe only whe n the Ru-33 is installed outside
the manhole.
Configur ing the Monitor t o S upport the Ru-33
Configur ing the TRITON+ to support the Telog Ru-33 requires
setting three paramete rs in Qstart when establishing a new location.
Refer to Create and Configure the Monitor Location in Chapter 6,
Configuration and Activation, for more information.
• Scan frequency This represents the interval at which the
monitor takes readings. These readings are available for
retrieval dir e c tly from the monitor, b ut ar e overwritten with
new data at each new interval. Therefore, ADS recommends
using the same interva l for both the scan r ate in the monitor and
the rate at which the Telog unit requests the data from the
monitor.
• Modbus ID This represents the Modbus slave ID that will be
used by the Ru-33 to indicate the specific monitor from which
to request the data. ADS recommends using 55 to identify the
TRITON+.
• Modbus Delay Response Thi s repr esents the amount of
time, in milliseconds, following the r e quest before the monitor
will retur n the requested Modbus output data to the Ru-33.
ADS recommends maintaini ng the 10 ms default setting.
Connecti ng t he Monitor to the Ru-33
Connecting the TRITON+ to the Ru-33 requires the ADS Triton-
Telog Comm Cable (ADS p/n 8000-0054-01). First, c onnect the
black, metal connector (with gray, plastic boot) from the comm
cable to the COMM + EXT PWR port on top of the TRITON+.