User Manual
D-1
A P P E N D I X D
System Configuration and Setup to
Support the Telog Ru-33
Using a battery-powered ADS
®
TRITON+
™
to support the Telog
®
Ru-33 Recording Telemetry Unit (RTU) requires some limited
configuration and setup. These procedures involve de s i gnating the
appropriate identification informa tion to enable recognition and
data exchange between the ADS and Telog units and using an
interface cable to connect the monitor to the RTU.
Note: The Telog Ru-33 must have fi rmwa re vers ion
3.300 or later to effectively request data from the
TRITON+ monitor.
Connecting the units togethe r primarily involves running the ADS
Triton-Telog Comm Cable from the battery-powered Ru-33 to the
battery-powered TRITON+ monitor. This applic a tion requires
installing both unit s in the manhole , a hazardo us environment.
However, ADS offers an optional extension cable (ADS p/n 8000-
0044-25) that allows you to install the Telog unit outside the
hazardous environment.
Note: The Telog Ru-33 does not possess IS (intrinsically-
safe) certification. Therefore, an installation involving
conne cting the Telog to the IS TRITON+ with both units
residing inside the manhole will not be considered
intrinsically safe. The installation will be considered