User Manual
Specifications A-13
XBUS (ADS p/n 8000-0427)
The XBUS
™
(External Modbus Interface unit) enables the
TRITON+ to receive power from an external AC power source and
supports Modbus communications.
Enclosure
Indoor/outdoor NE MA 4X (IP 66), PBT and
polycarbonate plastic with hinged cover
Dimensions
11.02 inches (280 mm) long x 7.49 inches (190
mm) wide x 5.03 inches (128 mm) deep
Weight
4.0 pounds (1.8 kg)
Operating/
Storage
Temperature
14° to 122° F (-10° to 50° C)
Power Supply
XP Power DNR60US24 (Refer to page A-12 for
specifications and details on t he power supply)
Power Input
85-264 VAC, 120-375 VDC; 47-63 Hz; 1. 10 A @
110 / 0.59 A @ 250 V AC
External Power/
Communications
Unit
ADS ExPAC (Refer to page A-11 for
specifications and details on t he ExPAC)
Connectors
Three cable glands for power cable, monitor
com municat i on c abl e, and RS-232/RS-485 cable,
NEMA4X/IP66 protec tion
Certification
Associated Apparatus certif i c ation for use with
approved equipment i n Zone 0 (Class I, Division
1, Groups C & D, i n the U.S.) and CSA Cl ass I,
Zone 0, II B
Sira 09ATE X2027X, IECEX SIR09.0020X, CSA
2013 2671180