User Manual
A-4 ADS TRITON+ Manual
represents the actual entities plus one d a te /time stamp. A
date/time stamp must be include d for each monitoring point
associated with the monitor.
Peak Combo
Sensor
Surface Combo
Sensor
Ultrasonic Dep th
Sensor
7 6 3
• Deduct 2 to 3 days from the total days of storage to compensate
for additional overhead in memory.
Following is the equation for calc ulating the total d a ys of storage
available in the monitor memory:
Data Storage Available in the Monitor (in Bytes)
=
Total Days of
Storage in
the Monitor
Memory
Number of
Readings
Taken per
Day for
One entity
X
Memory
Consumed for
One Entity
Reading
(Date/Time
Stamp)
X
Number of
Entities Logged
by the Moni tor
(based on the
selected
devices)
For example, recording data at a 15-minute sample rate using a
Peak Combo Sensor offers approximately 150 days of data storage.
851,968 bytes
=
158 days (150
when acc ounting for
overhead)
96 readings X 8 bytes X 7 entities (including
1 for the date/ tim e
stamp)