9-4 ADS TRITON+ Manual
XBUS
CABLE GLAND
RS-485 CABLE
Running RS-485 cable t hrough left cable gland i n XBUS
3. Loosen the screws on the ter minal connector.
4. Make sure the wires from the cable are stripped and separated.
5. Insert the nega tive wire (D-) from the cable into the negative
(-) terminal on the connect or (refer to the label adjacent to the
port), and tighten the terminal screw.
6. Insert the positive wire (D+) from the cable into the positive
(+) terminal on the connector, and tighten t he terminal screw.
7. Insert the ground wire (GND/SHL) from the cable into the
ground (
) terminal on the connector, and tighten t he terminal
screw.