User Manual

6-68 ADS TRITON+ Manual
Editing the XIO Device
Use the following infor mation to configur e the exter nally powered
XIO
device for the TRITON+ monitor for up to two analog
inputs, two analog outputs, two digital inputs and two digital
outputs:
XIO Device properties dialog
Note: There is a single XIO device limit per TRITON+.
Analog Input 1 (2) Configure the values in this section to sp e c ify
how to proce ss the inco ming analog signal from an analog device
connected to the XIO device.
Entity Sele c t the entity type c orre s pondi ng to the incoming
analog signal to ensure the correct unit of measure is applied to
the resulting data. The XIO device supports the following
analog inputs: depth (DEPTH_A1), velocity (VEL_A1),
quantity (Q_A1), temperature (TMP_A1), PH (PH_A1), as
well as a generic analog input (AI_A1). For example, choose
DEPTH_A1 to record incoming analog depth values to a
DEPTH_A1 entity which uses corresponding units of measure
in inches (or millimete rs).