Trigger Enter the s ampling tri gger thre s hold corresponding to
the selected Type for which you want the sampler activities to
initiate. For example, if Depth is the selected Type, enter the
corresp onding threshold value in inches (or millimeters) a t
which you want the monitor to pulse the sampler.
Samples per wakeup Select and enter or use the up/down
arrows to indicate the total, c onsecutive number of time s you
want the monitor to pulse the sampler within a single interval
each time the trigger condition is met.
Pulse Length Ente r the duratio n of the pul s e you want to
send to the sampler once sampling activities begin.
Data Optimization
Veloc ity Scrubbi n g Select this checkbox to ensure the
monitor confirms questionable or erroneous velocity readings
based on the historic a l r e la tionship between the ultrasonic
depth and velocity data for the location.
Automatic Pressure Calibration Select this checkbox to
ensure the monitor automatically c a librates questionable o r
erroneous pressure depth data to read consistently with the
ultrasonic depth da ta . The monitor will mainta in a running
average of the d ifference between the pressure depth and the
ultra s onic d epth data throughout the day. At midnight each
day, the monitor will apply a new offset based on the most
recent differ ence in the running average.
Combo Sensor to Use
Depth Select the combo sensor containing the depth sensor
you want to use for process es involving MLI and to generate
UNIDEPTH.
Velocity Select the combo senso r co ntaining the velocity
sensor you want to use for pr ocesses involving MLI and to
generate RAWVEL.