User Manual

Configura tion and Activation 6-59
outside a designated tolerance. The monitor calculates the
differences usi ng a running ave rage to minimize the p ossibi l ity of
anomalies that may trigger false alarms. The monitor also will log
an event when the d ifference in readings falls back below the
designated tolerance.
Enable Depth Select this checkbox to ensure the monitor
logs a CROSS_DEPTH event when the difference between at
least two of the depth readings at the monitoring point exceeds
the corres ponding tolerance.
Tolerance Enter the d ifference in depth measurements
beyond which you want the moni tor to log a cross c heck event .
Enable Velocity Select this checkbox to ensure the monitor
logs a CROSS_VELOCITY event when the difference between
at least two of the velocity readings at the mo nito ring point
exceeds the corresponding tolerance.
Tolerance Enter the d ifference in velocity measurements
beyond which you want the moni tor to log a cross c heck event .
Sampling
Edit the para meters in this section in order to manage water quality
sampling activities associated with a TRITON+.
Note: Water quality sampling cannot be performed on
externally powered TRITON+ monitors.
Type Select the triggering method you want to use to initiate
sampling operations from Depth, Flow Rate, or Flow
Total. Trigger values are based on readings returned from the
selected sensors in the Combo Sensor to Use section of the
MLI device. Choose Depth to initiate sampling based on
unidepth readings from the selected depth comb o sensor. Flow
Rate and Flow Total are instantaneous calculations based on
the selected depth and velocity senso r s in the Combo Sensor
to Use.