User Manual

Configura tion and Activation 6-35
Store D ata Se le c t this checkbox to ensure the monitor logs
the velocity data to memory. This option is selected by default.
Advanced Select this button to ac c e ss the Advanced
Parameters dialog.
Advanced P ar ameters dialog
Note: Modifications made to the Ad vanced Parameters
should be performed only by or under the guidance of
trained ADS personnel. Therefore, ADS r eco mmends
leaving the default settings (Auto) in place.
Modify the parameter s as necessary, and then select OK once
complete.
DAC Gain Choose the method for managing the
amplification of the analog signal that returns to the sensor:
Auto Select this op tion to allow the monitor to
automatically optimize the amplification of the signal.
Manual Select this option to e nter a fixed value
between -12db (lower amplification) and +12db (high
amplification) at which to amplify the signal.