Electronic Offset Enter the difference that exists between
the depth sens or re adings and the manually measured depth.
The default value is 0.
Deadband Select the distance from the sensor face within
which readings received will be ignored. The default value
(Auto) represents a percentage of the flow depth and should be
sufficient under normal circumstances. It a lso is the b e st option
for ha ndling shal l ow flows. However, consider another value
when the sensor readings reveal a potential problem with false
drop-outs. Choose a value for the deadband that is sufficiently
below the minimum depths that legitimate ly occur at the
monitoring point, yet above the confirmed drop-outs that have
been detected.
Store Data Select this checkbox to ensure the monitor logs
the depth data to memory. This option is selected by default.
Advanced Select this button to ac c e ss the Advanced
Parameters dialog.
Advanced P ar ameters dialog
Note: Modifications made to the Ad vanced Parameters
should be performed only by or under the guid ance of
trained ADS personnel. Therefore, ADS r eco mmends
leaving the default settings (Auto) in place.