User Manual

Introduction 1-13
Note: This is an excerpt from an agency-controlled
document for il l us tration purposes only. Changes to the
base controlled document require agency approval. For a
full-size copy of this drawing, please contact ADS and
request drawing number 8000BK0009-CERT.
Note: The Telog Ru-33 can be connected to the COMM
+ EXT PWR port on the TRITON+ monitor using the
Telog-Triton Comm Cable (ADS p/n 8000-0054-01).
However , this configuration has not been i ncluded on this
drawing because the Telog unit is not certified for IS
operation. Therefore, an installation involving this
configuration will not be considered intrinsically safe,
unless the Ru-33 is installed outside the hazardous area.
Cellular Modem Information a nd Compliance
Wireless telemetry is provided via a third-party, FCC - and carrier-
approved, commercial HSPA+/GSM or CDMA/EV-DO modem
inside the TRITON+ monit or and a corresponding antenna. Two
antenna opti ons are available thr ough ADS. However, customers
may o btain their own antennas to accommodate specific needs or
requirements. If a customer -supplied antenna is preferable or
required, consult an ADS or IETG, as applicable, representative
prior to installing the antenna to ensure it will adequately support
TRITON+ monitor communications and compliance.
European ATEX Hazardous Area Compliance
The following instructions apply to equipment covered by the
certificate numbered Sira 09ATEX2027X (TRITON+, U SB Serial
Interface, and Combo Sensors CSX Series). Reference European
ATEX Directive 94/9/EC, Annex II, 1.0. 6:
The equipment may be used with flammable gases and vapors
with apparatus groups IIA and IIB and with temperature classes
T3 (152
o
C), T4, T5, and T6.