34 Signal Connections
PCI-9524 ADLINK Technology Inc.
User’s Manual Copyright 2008
Floating Signal Sources
A floating signal source is not connected in any way to the build-
ings ground system. A device with an isolated output is a floating
signal source, such as optical isolator outputs, batteries, trans-
former outputs, load-cells and thermocouples.
3.2.2 Input Configurations
Differential input mode for transducer input channels
AI channels 0 to 3 are dedicated to connecting to load-cell trans-
ducers in differential mode.
A load-cell is comprised of four resistive strain-gauges connected
in Wheatstone bridge form, and is inherently a floating differential
output device. Since a load-cell transducer is a passive device, it
requires voltage excitation in order to transform the resistive
change into electrical signals. A typical four-wire connection is
shown on Figure 3-4.
Figure 3-4: Connecting to a four-terminal load-cell transducer
using a four-wire connection
+Vex
+Vo
-Vex
-Vo
AIn+
AIn-
VEXn+
VEX_SENn+
VEXn-
VEX_SENn-
NC
NC
Load-cell Transducer PCI-9524
It is recommended to enable the remote-sense function, and loop-
back the VEXn+/- to VEX_SENn+/- on the terminal board you're