Instruction manual

48
10 GB Communication
Version
1.0
3
& Control Card Instruction Manual
SETCLOCKCONTROL instruction
Byte
7
6
5
4
3
2
1
0
1
-
2
Opcode
0x0
102
3
-
4
Length
0x0001
5
0
0
0
SS
AA
AS
0
0
Description:
Select ADC Clock and Sample source. (Use the SAVESETTINGS
instruction to save changes to the EEPROM.)
AS: 0 – Internal AD Clock Source
1 – External AD Clock Source
AA: 0 – Normal External Clock Mode
1 – Auto External Clock Mode
AS: 0 – Internal Sample Source
1 – External Sample Source
SETCLOCKENABLE instruction
Byte
7
6
5
4
3
2
1
0
1
-
2
Opcode
0x0
103
3
-
4
Length
0x0001
5
0
0
0
0
ES
CS
EC
CC
Description:
Enable or disable clock outputs. (Use the SAVESETTINGS instruction to
save changes to the EEPROM.)
ES: 0 – Disable sample output of the EIO connector
1 – Enable sample output of the EIO connector
CS: 0 – Disable sample output of the CONTROL connector
1 – Enable sample output of the CONTROL connector
EC: 0 – Disable clock output of the EIO connector
1 – Enable clock output of the EIO connector
CC: 0 – Disable clock output of the CONTROL connector
1 – Enable clock output of the CONTROL connector