Instruction manual

25
10 GB Communication
Version
1.0
3
& Control Card Instruction Manual
4.7 Parallel Data Interface
The 10 GB C&C Card has a two-way parallel port, through which data
can directly be loaded from IP to the slave device or from the slave device to
IP. Several slave devices can be connected to the bus. The Parallel Data
Interface (PDI) has an 8-bit bi-directional data bus (DATA 0-7) and control
signals: Clock (CLK), Transfer Enable (TE), Read/Write# (RW), IRQ#, WAIT#
and RESET#.
Pin
Description
Direction
IO Standard
Impedance
1
GND
-
-
-
2
DATA 0
Bi
-
directional
3.3 V CMOS
2.2
kΩ pull
up
3
DATA 1
Bi
-
directional
3.3 V CMOS
2.2
kΩ pull
up
4
DATA 2
Bi
-
directional
3.3 V CMOS
2.2
kΩ pull
up
5
DATA 3
Bi
-
directional
3.3 V CMOS
2.2
kΩ pull
up
6
VDD (3.3 V)
-
-
-
7
IRQ
#
Input
3.3 V CMOS
2.2
kΩ pull
up
8
DATA 4
Bi
-
directional
3.3 V CMOS
2.2
kΩ pull
up
9
DATA 5
Bi
-
directional
3.3 V CMOS
2.2
kΩ pull
up
10
DATA 6
Bi
-
directional
3.3 V CMOS
2.2
kΩ pull
up
11
DATA 7
Bi
-
directional
3.3 V CMOS
2.2
kΩ pull
up
12
WAIT
#
Input
3.3 V CMOS
2.2
kΩ pull
up
13
RESET
#
Output
3.3 V CMOS
2.2
kΩ pull
down
14
RW
Output
3.3 V CMOS
-
15
TE
Output
3.3 V CMOS
-
16
CLK
Output
3.3
V CMOS
27 Ω
serial
PDI data rate: 125 kB/s
A Read or a Write operation on the Parallel Data Interface can be performed
using PDIREAD or PDIWRITE instructions. (Details about the PDI can be
found in the GEC Instruction Manual Section 4.7)