User Manual
________________________________________________________________________________________________________________
76
trim,thenyoucan pressEnter or Donetodisplay theOutput Trim
interface(asshowninFigure11.12).
3. In the Output Trim interface, press Enteror Trim to execute the
outputpointadjustment.Thetrimvaluecanbefetchedbypressing
Fetch,oritcanbeinputtedmanually.
4.
IfOutputTrimissuccessfullyexecuted,thecalibratorwillgoesback
totheOutputTrimpointselectioninterface.Youcanrepeatsteps2
and3tostartanotherOutputTrim.
Figure11.12HARTOutputTrim