70
abortedimmediatelyandshowtheHARTdevicelist.Otherwisecalibra torwillcontinuesearchingfromaddress1to
15,andautomaticallystoppollingataddress15.Intheperiodofpolling,youcanpressStoptostoppolling.Ifmore
than one hart device is found, calibrator will show the HART devices list,
otherwise calibrat or will prompt you
whethertopollHARTdeviceagain.Amaximumof15HARTdevices
canbesupported.
3. Press Enter or Select to Select a HART device from the list, the
calibrator will connect to the selected HART device, and the HART
deviceinformationwillbedisplayed
(AsshowninFigure11.3).
4. PressEnter or Done , then the calibrator will go back to the main
screeninBasicMode,theHARTcommunicationinterfacedisplayed
inthemeasurementdisplayarea,andtheHARTiconwillbeshown
inthestatusbar(Figure11.4).TopressEsckey
ortheBacksoftkey,
youcanreturntotheHARTdevicelisttoselectanotherHARTdevice
togetthedeviceinformation.
WhentheHARTCommunicationInterfaceisactive, theProcess, Service
and Setup softkeyare available. You can press the softkey to access to
thecorre spondingHARToperations.
11.3 HART Process Variables Update
Whenthe HART Communication Interface is active,
you can
pressProcesstoshow HART Process Variables interface (Figure 11.5). In
Figure11.5HARTProcessVariables