Pump the pressure to your target value until the pressure is stable (To avoid damaging the pressure module from
overpressure,neverapplypressureabovetheratedmaximumprintedonthepressuremodule).Press EnterortheStart
tostarttesting.
Whentheteststarts,thetesttimeescapessecondbysecond.Theleak
testwillbestoppedwhenthetesttimeiszero.
ThenthetestresultswillappearintheReportgroupbox.
Caution:Ifthepressureofthepressuremoduleisoverrange,thesemightbeaproblemfetinganavailableresult.
9.4.6 Unit Converters (Temperature, Pressure, Mass Flow and Volume Flow)
TheutilitiesincludefourUnitConverters:Temperature,Pressure,MassFlow
andVolumeFlow.
Ifthevalueofanyeditboxischanged,otherunits’valueswillbechangedautomatically.
9.5 System Settings
9.5.1 Alarm Settings
IntheAlarmSettingspage,thereisabeepONorOFFoptionincaseofMeasurementoverrange,SimulateTransmitter
overrangeandPressureModuleoverrange.
Cautions:
Toprotectthecalibrator,
neverapplymor ethan30Vbetweenanytwoelectricaljacks
Toavoidpossibledamagefr omtheconnectiontotheoutputsignalloop,scalethesimulationoutputsignals.
Toprotecttheexternalpressuremodule,donotoperatethedeviceoveritspressurerange.
9.5.2 Screen Settings
Youcansetthebrightnessin
thispage.Lowerbrightnessvaluescansavemorebatterypower.
TheAutoBrightnessOffdropdownlistincludesthetimeoutperiodofthebacklight.Thebacklightswillbeturnedoff