49
9.4.3 Custom RTDs Library
ThecalibratorcanmeasuretemperaturewithmostRTDs.ButsometimesthereisaneedtosupportnonstandardRTDs
andStandardPRTs(standardplatinumresistancethermometers).
Thecalibratorallowsyoutoinputthecustomizedvalues,soyoucanmeasuretemperatureusingeverykindofRTDsand
StandardPRTs.
How to create a custom industry RTD?
To get high
precision, IEC and ASTM standards specify a more complex polynomial that finetunes the
resistancetemperaturerelationship.Oneformofthatequationisgivenhereasbelow.TheindustryRTDsalsospecify
valuesfortheconstantsR0,A,B,C(TheCconstantisonlyusedfortemperaturelessthan
0).

1 
 100
Forexample,forthecommonPt385RTDwith100Ωresistanceat0°C,thevaluesare:
100Ω
A 3.9083 10

B5.77510

C 4.183 10

How to create a standard PRT?
The Standard Platinum Resistance Thermometer (SPRT) Calibration Laboratory of the NIST ThermometryGroup can
distinguish among the ITS90 from the argon triple point (Ar TP, –189.3442 °C) to the silver freezing point (Ag FP,
961.78°C)forthecalibrationofSPRTs.
TheITS90definestemperaturethroughasetof
specifiedthermometric fixedpoints, interpolationinstruments, and