User Manual
________________________________________________________________________________________________________________
43
7. Usingexternalspecialpressuregeneratortoadjustthepressurevalue.
8.3.13 Thermocouples Simulation
Refer to Table 5‐3 for dataabout thermocouple types included in the
calibrator.
Proceedasfollowstosimulateathermocouple:
1. Ifsourceisactive,press once,otherwisepress twice.
2. You can pressthe navigationkeysUp/Downtoselect the
first item
fromthepop‐uplist,andthenpressEnterorDonetoswitchsource
to thermocouples simulation or press Options to show
thermocouples simulation setup interface. In the thermocouple
simulation setupinterf ace,you can set the thermocouples sensor
type, tempera ture units( , K,℃ ) and reference junction℉
compensationmethods.
3.
In thermocouple simulating, you can press Options to show
thermocouple simulation setup interface and set the simulation
parameters.
4. Refertothesections8.3.1,8.3.2,8.3.4or8.3.5toselectamethod
tochangethesourcevalueaccordingtoyourneed.
8.3.14 RTD Simulation
Refer toTable5‐4forinformationaboutRTD
(ResistanceTemperature
Figure8.17ThermocoupleSimulation