switchsourcetocurrent.
4. Whensourcingcurrentwithinternalpower,the24VDClooppowerstatesmustremainasON.Ifthe24VDCpower
stateisOFF,itwillbeautomaticallyturnedbackONstatesafterswitchingsourcetocurrentwithinternalpower.
5. Refer to the sections8.3.1,8.3.2, 8.3.3,
8.3.4or 8.3.5,select amethod to change the sourcevalue accordingto
yourneed.
8.3.12 Pressure source (only available for pressure
module)
Proceedasfollowstopressuresource:
1. Figure8.16showsthepressuresourceconnections.
2. Connectthepressuremodule.
3. If source is active, press once, otherwise press
twice.
4. Press the navigation keys Up/Down
to select the first
item from the pop‐up list, then press Enter orDone to
changethesourcetopressure.
5. YoucanpressUnits toviewthepressureunitslistwhich
are supported by the calibrat or. You can change the
pressureunit.
6. You can press Module to view the
information about
thepressuremodule.
Figure8.16PressureSource