Figure8.13RampingConfiguration
mainscreen.
4. PressAutotobeginautomaticstepping,andthesoftkeylabelwillchangetoPause.
5. PressPausetopauseautomaticstepping.ThesoftkeylabelwillchangetoResumeatthesametime.Youcanpress
Resumetocontinueautomaticstepping.
Ifyouwanttoexitautomaticstepping,pressEscor
Exit.
8.3.5 Ramping the source value
In the Basic Mode, the calibrator offersthe source value ramping.
Rampingcansmoothlyandcontinuouslyincreaseordecreasethesource
value. The size ofsteps is determined by thestart value, end value and
theramptime.Beforeramping,youneedtoconfiguretherampsettings.
Each source item
has its own ramp settings. When source is active and
source item is not pulse, frequency and pressure, you can press Ramp
toview the ramp setup interface. Figure 8.13 and Figure 8.14 are
examplesoframpsettingsandtherampresultsofsettings.
The Start value cannot equal the End
value, and their values cannot
exceedthesouringitem’soutputrange.
Note:Theramptimeisasinglerisingtime,alsoisasinglefallingtime,so
theramprisingprocessandthefallingprocessissymmetric.Inthecase
of fixedstart value and end value, the longer ramp time it
takes the
slowerramp and smallerstepssize itwillbe.Therange of ramp timeis