8.2.10 RTD Measurement (Temperature)
The calibrator supports11 common RTDs shown in Table 54. In
addition, calibrator accepts 10custom RTDs.to add a custom RTD
please refer to 9.4.3. The calibrator accepts RTD measurement
inputs in twowire, threewire, or fourwire connections as shown
Figure 8.10. The fourwireconnection can provide the highest
measurementprecision.
When using RTD input to measure temperature, proceed as
follows:
1. If measurement is active, press once, otherwise press
twice.
2. Press the navigation keys Up/Down to select the second item
fromthepopuplist,andthenpressEnterorDonetochangethe
measurement or press Options
to view the setup interface. In
setup interface, you can set the RTD sensor type, connections
andtemperatureunits(,K, ).
3. WhenRTDmeasuring,youcanpressOptionstoviewthesetup
interfaceandsetthemeasurementsettings.
4. To get higher measurement precision, you can zero measured
value
duringresistancemeasurement.
Figure8.10RTDMeasurement