User Manual
________________________________________________________________________________________________________________
13
Simulating Thermocouple Source&using Thermocouple to Measure Temperature Accuracy. (Working temperature:
20℃±5℃)
Table5‐3
Thermocouple Measurement and Source accuracy
Measure and
Simulate
Standard
Temperature Range (℃)
Accuracy (℃)
Measure Source
S IEC584‐50to1768
‐50to400 1.0 1.1
400to1000 0.6 0.6
1000to1768 0.7 0.8
R IEC584‐50to1768
‐50to200 1.4 1.4
200to500 0.6 0.6
500to1768 0.6 0.7
B IEC584 0to1820
50to450 3.8 3.8
450to800 0.9 0.9
800to1820 0.7 0.7
K IEC584‐270to1372
‐250to‐200 1.0 1.1
‐200to‐100 0.4 0.5
‐100to600 0.3 0.3
600to1372 0.4 0.5