Manual
Pressure / Process Calibration EquipmentAdditel Datasheet 2014 / 2015
Corporate Headquarters
22865 Savi Ranch Parkway Ste F
Yorba Linda, CA 92887, USA
Salt Lake City Office
802 Bamberger Dr, Suite C
American Fork, UT 84003, USA
Phone: 714-998-6899
Fax: 714-998-6999
Email: sales@additel.com
Phone: 801-756-5570
Fax: 714-998-6999
Email: sales@additel.com
Environmental
Compensated Temperature: 14°F to 122°F(-10°C to
50
°
C )
Operating Temperature: 14°F to 122°F(-10C to 50C )
Storage Temperature: -4°F to 158°F (-20°C to 70°C )
Humidity: <95%
Pressure Connections
1/4NPT male, 1/4BSP male, M20X1.5 male
Other Connections available per request
Enclosure
Case material: SS
Wetted parts: 316L SS
Dimension: Ø1.3" x 4.84" (Ø33mm x 123mm)
Weight: 0.99 lb (0.4kg)
Warranty: 1 Year
Absolute Pressure
P/N
Pressure Range
Media Accuracy(%FS)
Burst
Pressure
(psi) (bar)
AP5 5 0.35 G 0.1 3×
AP10 10 0.7 G 0.1 3×
AP15 15 1.0 G 0.1 3×
AP30 30 2.0 G 0.1 3×
AP50 50 3.5 G 0.1 3×
AP100 100 7.0 G , L 0.1 3×
AP300 300 20 G , L 0.1 3×
AP500 500 35 G , L 0.1 3×
AP1K 1,000 70 G , L 0.1 3×
AP3K 3,000 200 G , L 0.1 3×
AP5K 5,000 350 G , L 0.1 3×
SPECIFICATIONS
ORDERING INFORMATION
Model Number
Model
Range type:
PSI - range by psi
BAR - range by bar
Pressure port type:
N – 1/4NPT male
N2 - 1/2NPT male
B -1/4BSP male
M- M20 X 1.5 male
ADT160 GP300 PSI N02
Pressure range P/N:
See pressure range table
Accuracy:
02-0.025% of full scale
05-0.05% of full scale
10-0.1% of full scale
Accessories included
NIST Traceable Calibration Certicate
Optional Accessories
Model
number
Description Picture
9060 Pressure module connection cable
Differential Pressure
P/N
Pressure Range
Media
Accuracy
(%FS)
Burst
Pressure
Static
Pressure
Range
(inH
2
0) (mbar)
DP1 ±1 ±2.5 G 0.05 100× ±10 psi
DP2 ±2 ±5.0 G 0.05 100× ±10 psi
DP5 ±5 ±10 G 0.05 50× ±10 psi
DP10 ±10 ±25 G 0.05 20× ±10 psi
DP20 ±20 ±50 G 0.05 20× ±10 psi
DP30 ±30 ±75 G 0.05 20× ±10 psi
DP50 ±50 ±160 G 0.05 3× ±10 psi
DP150 ±150 ±350 G 0.025 3× 50 psi
DP300 ±300 ±700 G 0.025 3× 50 psi
Additel 223A with ADT160 Pressure Module
* One year accuracy (including yearly stability) except DP1 and DP2
range which is 0.05%FS calibration accuracy and 0.05%FS yearly
stability.