Technical information

C:
\
UFFICIO T
ECNICO
\
MANUALI
\
ML2
00
\
ML200 INGLESE
\
68
TRASFORMATION CONVERTER FROM SEPARATE VERSION TO
COMPACT VERSION
Cover
Fixing screw
Fixing plate
O-Ring 2056
O-Ring
Grower Ø 6
Nut Ø 6
SENSORE
SENSORE
- Remove the connector
- loosen the cable gland and
remove the cable from
connection box
cable gland
- Insert the connector
of compact version
nei connettori femmina
(polarized connector)
- Remove the cable gland
- Insert the corcks PG 9 inclusive in
the kit and tighten the corks with the
appropriate nuts
cover of connector
box
Wall bracket
- Disconnect the cables from
terminal
- Remove the cable from
the converter
- Unscrew the fixing flie
- Unscrew the fixing screw
Fixing flie
- Uncrew the nut and unthread
the fixing plate of separate
version
- Assemble the fixing plate
of compact version like
visualized in pic. 1
- Remuve the cover of
separate version
- Insert the cable in the hole
of box converter
- Assemble the box converter to
connector box
- Tighten the fixing screw
Fixing screw
1
2
3
4
5
N.B.:
For the connections of cables to the converter see section: "Electrical connections" of operative manual
SENSORE
CONVERTITORI
ML 190
ML 200
ML 201
ML 202
SENSORI
MS 500
MS 1000
MS 2400
MS 2500
MS 3700
MS 3770
Rear
converter
IT019
1 2 3 4 5 6 7 8 9 10
11 12 13 14 15 16 17 18 19 20