Technical information
C:
\
UFFICIO T
ECNICO
\
MANUALI
\
ML2
00
\
ML200 INGLESE
\
67
APPENDIX 1
TRASFORMATION CONVERTER FROM COMPACT VERSION TO
SEPARATE VERSION
2
ISOIL
Connection box
section
1 2 3 4 5 6 7 8 9 10
11 12 13 14 15 16 17 18 19 20
Cover
Power supply
- Disconnect the cable from
terminal
- Unthread the cables from converter
ISOIL
O-Ring
Grower Ø 6
6 MA nuts
- Uncrew the 6MA nuts
and remuve the fixing
plate for compact
version
Rear
converter
ISOIL
Fixing plate
O-Ring 2056
- Assemble the fixing plate for
separate version like
in picture aside
Grower Ø 6
ISOIL
Fixing screw (tighten strongly)
Wall bracket
Fixing flie
- Tighten the
fixing screw
- Screw the fixing
flie
Nut of cable
gland
Coil connector
(2 cable)
Electrode connector
(3 cavi)
- Insert the connectors in the
connectors assemblate to
the connectors plate
- Insert the cables in the connectors
box and tighten nuts and cable
gland
O-RING 155
cover of
connection box
- Remuve the corcks
3
- Insert the cable in the
connector box like in
picture aside
- Insert the l'O-Ring
in the appropriate seating
- Assemble the cover
to the connection box
by screwing the
fixing screw
4
5
6
- Make the electric connections following
the indication listed in section:
"Electrical connection" of meter
operative manual
- Remuve the fixing screw and
separate the sensor from converter
by remuving the connectors
1
6 MA nuts