Technical information
C:
\
UFFICIO TECNICO
\
MANUALI
\
ML2
00
\
ML200 INGLESE
\
6
Flow direction
GENERAL INFORMATION ON THE INSTALLATION
RECOMMENDATION FOR MS 500-1000-2400-2500-3700-3770
The hooks are project exclusively
for the weight of the instrument
Recommended lifting
Centring cylinders
(before support the sensor,
making sure that the centring
cylinders be in the middle of
the sensor)
Method of upheaval recommended for all the
sensors equipped from eyebolt.
For sensor MS 1000 we recommend the use of
centring cylinders (available on request)
Before tighten the nuts make sure that the
ending of the sensor are coaxial to those of the
pipe
Before install the sensor locate the direction of
the liquid in the piping
The sign of the flow rate is positive, when the flow direction
it’s (from – to +) as printed on the tag plate.
If after the installation, for plant request becomes
necessary reverse the sign of the flow, it's enough reverse
the sign of the coefficient KA
ISOIL
//
//
ISOIL
?
ISOIL
- FLOW DIRECTION +