(POS. 53) Data logging time interval set [INTERV.(h)= X]
Sampling time interval for the data logging function and their printing. The allowed values are: 1, 2, 3, 6, 8, 12, 24, 48
hours.
(POS. 54) Date and time set [? = dd/mm/yy hh:mm]
Date and time set. If ME 220 optional module is present, then the time setting is kept also when the power supply is
off, otherwise it is frozen till the power supply is back. For example, if the power supply has been off for one hour,
when switched on the instrument will be one hour late. The calendar is valid till year 2091.
N.B.: date and time are visualized on display only if the data logger is enable.
(POS. 55) Logged data display [DISPLAY DATA]
Displaying of the data stored in the data logger. With this function it is possible to scroll through the gathered data.
(POS. 56) Logged data cancel [CLEAR DATA]
Logged data cancel function. With this function the storage memory is completely emptied.
MENU 10.DIAGNOSTIC
(POS. 57) “Autotest” function enable [SELF TEST]
Meter autotest function. This function stops the normal functions of the meter and performs a complete test cycle on
the measure input circuits and on the excitation generator. To activate this function, after select it, push key , at
the question: “EXECUTE?” push for more 1 second key (for more 1 second) for start autotest, or any other key for
delete operation. The result of the test is shown on the display. At the end of operation will have visualized one of
visualization page. This function is automatically performed when switching the device on.