Technical information
C:
\
UFFICIO T
ECNICO
\
MANUALI
\
ML2
00
\
ML200 INGLESE
\
57
MENU 8.DISPLAY
(POS.47) Totalisers number of decimals set [TOT.DECIMALS= X]
Totalisers number of decimal figures set. The allowed values are from 0 to 3.
(POS.49) Language [LANGUAGE= XX]
Choice of the layout language. There are 4 languages available: E = english, I = Italian, F = French,
S = Spanish.
(POS. 50) Display refresh frequency set [DISP.FR(Hz)= X]
Set of the frequency the data on the display are refreshed at. This parameter effects only the display layout and not the
response time of the meter itself. The possible choices are: 0.1 – 0.2 – 0.5 – 1, 2, 5 e 10 Hz.
(POS. 51) Display contrast set [CONTRAST= X]
Display visual contrast set. This function comes into operation only when leaving the function itself. The allowed
values are from 0 to 15. The contrast can change in relation to ambient temperature, for this reason the
contrast set can be change from visualization pages (pag. 30-31 fig. A, B, C, D) pushing the key for
8 second or more. In this way the contrast set that will be visualized at release of the key.
MENU 9.DATA LOGGER
(POS. 52) Automatic data logging enable [ACQUISITION= ON/OFF]
Automatic data logging enable for volume values. The data are sampled every time interval set on the following
function.