Technical information

C:
\
UFFICIO T
ECNICO
\
MANUALI
\
ML2
00
\
ML200 INGLESE
\
54
(POS. 42) Choice of the function and the range of current output n.1[OUT.mA1=X÷XX ±]
Choice of the function and the range of current output N.1. The current output N.1 is optional and it is mounted on the
main board. There are three field to modify for this function:
?? Scale zero: 4 or 0 mA
?? Full scale: 20 or 22 mA
?? Field: + = positive, - = negative, ? = both, -0+ = central zero scale
The values corresponding to the scale points are shown in the follwing chart:
CURRENT VALUES IN mA ASSOCIATE TO THE % VALUE OF FULL SCALE
REVERSE FLOW VALUE ZERO DIRECT FLOW VALUE POSSIBLE FIELD
? -110% -100% 0% +100% ?+110%
OutmA = 0 ÷ 20 + 0 0 0 20 20
OutmA = 0 ÷ 22 + 0 0 0 20 22
OutmA = 4 ÷ 20 + 4 4 4 20 20
OutmA = 4 ÷ 22 + 4 4 4 20 21.6
OutmA = 0 ÷ 20 - 20 20 0 0 0
OutmA = 0 ÷ 22 - 22 20 0 0 0
OutmA = 4 ÷ 20 - 20 20 4 4 4
OutmA = 4 ÷ 22 - 21.6 20 4 4 4
OutmA = 0 ÷ 20 ± 20 20 0 20 20
OutmA = 0 ÷ 22 ± 22 20 0 20 22
OutmA = 4 ÷ 20 ± 20 20 4 20 20
OutmA = 4 ÷ 22 ± 21.6 20 4 20 21.6
OutmA = 0 ÷ 20 0+ 0 0 10 20 20
OutmA = 0 ÷ 22 0+ 0 1 11 21 22
OutmA = 4 ÷ 20 0+ 4 4 12 20 20
OutmA = 4 ÷ 22 0+ 4 4.8 12.8 20.8 21.6
In hardware alarm conditions (interrupt coils, empty pipe, measure error) the current value is programmed by
the function “mA VALL. FAULT” (page 49 pos. 28) and it is expressed as percentage of a fixed current range,
where: 0% = 0 mA e 110% = 22 mA.
Example 1 page 52
Example 2 page 52