that the liquid is perfectly staying still. Even very small movement of the liquid may effect the result of this function,
and, consequently, the accuracy of the system.
(POS. 36) Range change external command enable [RANGE CHANGE= ON/OFF]
Range change external command enable. When this function is enabled, applying a voltage on the on/off input
terminals the meter switches to the second measuring range (Fs2).
N.B.: the autorange doesn’t allow to use the manual change range (see pos. 22)
(POS. 37) Batch start/stop external command enable [BATCH= ON/OFF]
Batch start/stop external command enable. (see “BATCH FUNCTIONS”)
MENU 6.OUTPUT
(POS. 38) Choice of the function corresponding to on/off output 1 [OUT1= XXXXXX]
Choice of the function corresponding to digital Output 1. The functions are listed in the table at page 53.
(POS. 39) Choice of the function corresponding to on/off output 2 [OUT2= XXXXXX]
Choice of the function corresponding to digital Output 2. The functions are listed in the table at page 53.
(POS. 40) Choice of the function corresponding to on/off output 3 [OUT3= XXXXXX]
Choice of the function corresponding to digital Output 3. The functions are listed in the table at page 53.
Attention: Output 3 is Optional and is mounted on an optional add on module.