FUNCTION MENU
MENU 1.SENSOR
(POS. 1) Nominal diameter Sensor [ND= XXXX]
Sensor nominal diameter. The value shown in the pate on the sensor must be entered.
Its value has to be within the range from 0 to 3000 mm.
IMPORTANT: if you want to know the speed of the liquid passing through the sensor, set this parameter at 0. The
instrument will show the liquid speed expressed in meter per second (m/s). Attention: in this case all the totalised
values will be expressed in metres (m) and will then be without any meaning
(POS. 2) Coefficient KA [KA= ±X.XXXX]
Coefficient KA set. This parameter is calculated during the sensor calibration and stands for a signal amplification value.
It has to be set at the value shown in the plate on the sensor.
MENU 1.SENSOR
(POS. 5) Speed rate [S. RATE (Hz)= XXXX)
Menu visualized on the converter ( from 1 to 11)
Synthetic description of the function
Identification number of the function (not
visualized on display)
Converter request