Technical information
C:
\
UFFICIO TECNICO
\
MANUALI
\
ML2
00
\
ML200 INGLESE
\
24
ELECTRICAL CONNECTIONS SENSOR
-
CONVERTER
Sudden movements of the electrodes cable, can cause
noises on measure
CONVERTER ML 200
–
INTERNAL REAR VIEW
-
SEPARATE VERSION
Max length of cable: 20 mt
PRE-AMPLIFER VERSION
Max length of cable: 500 mt
Cable C014
From pin 1 to pin 5:
cable connected
to the sensor
1
2
3
4
11
12
13
1 234 5 678 9
10
Green (electrode 1)
Black (electrode 2)
White/black (common)
Red (coil 1)
Black (coil 2)
White
Black
Orange
Red
Blu
Electrode 1 (connecting terminal 1 M1)
Electrode 2 (connecting terminal 2 M1)
Common (connecting terminal 3 M1)
Shield (connecting terminal 4 M1)
Common (connecting terminal 11 M1)
Coil 1 (connecting terminal 12 M1)
Coil 2 (connecting terminal 13 M1)
Green
Black
Shield
White
Shield
Red
Black
Black
From pin 6 to pin 10:
cable connected
to the converter
Electrodes cable
Coil cable
1
2
3
4
11
12
13
Electrode 1
Electrode 2
Common
Shield
Shield
Coil 1
Coil 2
C013
C012
1 2 3 4 5 6 7 8 9 10
11 12 13 14 15 16 17 18 19 20
Dangerous voltage:
- 60Vdc max
- 250 V max commutation
coil's circuit
Terminal block M1