e-sphyg 2 TM 9002 Automatic Sphygmomanometer User’s Manual Inspected in the U.S.A. Made in Japan tel: 631-273-9600, 1-800-232-2670 fax: 631-273-9659 ADC 55 Commerce Drive Hauppauge, NY 11788 ADC (UK) Ltd. Unit 6, PO14 1TH United Kingdom IB p/n 93-9002-00 rev 3 www.adctoday.com email: info@adctoday.
A Special Thank You... ® Thank you for choosing an ADC blood pressure instrument. We're proud of the care and quality that goes into the manufacture of each and every item that bears our name. Only the finest materials are used to assure you of a timeless instrument designed for optimum performance. You'll quickly appreciate the results, for you now own one of the finest sphygmomanometers that money can buy.
ADC® 9002 e-sphygTM 2 Limited Warranty American Diagnostic Corporation (ADC®) warrants its products against defects in materials and workmanship under normal use and service as follows: Table of Contents Introduction and Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 1. Warranty service extends to the original retail purchaser only and commences with the date of delivery. 2. The entire sphygmomanometer is warranted for one year. Warnings and Precautions . . . . . . . . . . . . .
Introduction and Intended Use The 9002 digital ADC® e-sphygTM 2 sphygmomanometer is a digital sphygmomanometer intended for measurement of systolic and diastolic blood pressure and pulse rate in adult patients, i.e., age 12 and above. This product is not designed for neonatal use. Inaccurate readings may result if it is used on a child's arm.
Mobile Base Assembly Parts Warnings and Precautions Warning • Do not use the 9002 in an explosive environment such as where flammable anesthetics exist or inside an oxygen chamber with strong electrostatic and electromagnetic fields e.g., mobile phones. 10 9 10B Warning • Use of this instrument on patients under dialysis therapy or on anticoagulant, antiplatelets, or steroids could cause internal bleeding.
9002M - Mobile Unit Parts & Tools Product Features DUAL DISPLAY COLUMN LCD and SEGMENT LCD (Simulated Mercury Column) The pressure can be read in the column LCD height and as digital value in the segment LCD. DUAL MEASUREMENT MODE "AUTO" and "MANUAL" Modes of Measurement Determination of blood pressure can be made either automatically by the oscillometric method or manually by the auscultatory method using a stethoscope. MODE AUTO MANUAL DEFLATION RATE MEMORY Maintained at 4.0 to 4.
9002W - Wall Unit Parts & Tools Name and function of each part MAIN UNIT ADC Item Item no. 1. e-sphygTM 2 9002 2. 6mm Allen Wrench 752M-116 3. Wall Bracket 952-103 4. Wall Basket with (3) screws / (3) mollys 5. Coiled tubing 885 6. Mounting Bolt/Washer (1) 972-101 7. Connector 9002-2 8. Connector, Male 891M 9. Connector, Female 891F 10. AdcuffTM & Bladder 850F 11. (2) Screws (for wall bracket) 952-104 12.
Name and function of each part Care and Maintenance Cuff Sizes CHILD (small adult) Arm circumferences of 7.48”-10.71” (19.0-27.2cm) ADULT Arm circumferences of 10.28”-16.1” (26.1-40.9cm) LARGE ADULT Arm circumferences of 13.5”-20.04” (34.3-50.9cm) LUER CONNECTOR (Male Luer to Female Bladder Tube) BLADDER TUBE (Female Luer Connection) COILED TUBING Because the device includes precision parts, care should be taken to avoid extreme temperature variations, humidity, shock, dust, and direct sunlight.
Error Displays and Troubleshooting Preparation for Use Connecting and Using the AC Adapter (All models) ERROR SYMBOL/SYMPTOM CAUSE REMEDY Measurement result cannot be recalled or is not stored. Blood pressure was taken in “MANUAL” mode. Measurement result is saved only when it is taken in “AUTO” mode. Cuff inflation does not start. Battery is exhausted. AC INDICATOR is lit even when 9002 is turned off. The column LCD is dim or brightness is not steady. The battery is inserted.
Error Displays and Troubleshooting BATTERY MARK flashes in the segment LCD when the battery charge is low. Only a few more measurements remain before the battery runs down. ERROR SYMBOL/SYMPTOM The battery needs to be recharged when BATTERY MARK is displayed. Measurement cannot be made. NOTE: The rechargeable batteries may not be completely full after their initial charge or when they have not been used for an extended period of time.
Error Displays and Troubleshooting ERROR SYMBOL/SYMPTOM Nothing is being displayed when MAIN POWER SWITCH is pressed down in the ON position. How to Assemble Wall Model CAUSE REMEDY OVER-PRESSURIZATION: The cuff was inflated to the maximum pressure because of movement of body, etc. Make sure that the patient stays still during measurement. MEASUREMENT ERROR: Measurement could not be made because of moving or talking during measurement.
How to Assemble Mobile Model M e m o ry Function Measured result obtained in "AUTO" mode is automatically saved in the memory and can be recalled until next measurement is taken. The result is not saved when the measurement is performed in "MANUAL" mode. 1. With base on its side connect upright outer pole to base using large (6mm) allen bolt. Tighten with large (6mm) allen key provided. Tighten hardware as securely as p o s s ible.
Cuff Connections Apply the cuff and stethoscope to the patient's arm in the same manner as the usual auscultatory method. Press START/STOP button. The cuff is inflated automatically after zero settings. Pressurization stops when pressure reaches the preset pressure value and pressure starts to descend. Deflation rate is displayed at the lower right corner of the segment LCD. NOTE: Press START/STOP button whenever the measurement should be interrupted. The device will rapidly exhaust air from the cuff.
Apply the cuff to patient's arm. Choose the right size of cuff for your patient. NOTE: Only these 3 cuffs may be used in auto mode. ARM CIRCUMFERENCE CUFF SIZE 10.28 to 16.1 inches (26.1 to 40.9 cm) ADULT 13.5 to 20.04 inches (34.3 to 50.9 cm) LARGE ADULT CHILD 7.48 to 10.71 inches (19.0 to 27.2 cm) The lower edge of the cuff should be about 1” (2 to 3 cm) above the inner elbow and BLADDER TUBE should be over the inner part of the arm, positioning the ARTERY MARK of the cuff over the brachial art e ry.