Guidance and manufacturer’s declaration - electromagnetic immunity - for all MEDICAL ELECTRICAL
EQUIPMENT and MEDICAL ELECTRICAL SYSTEMS
Table 2: Guidance and manufacturer’s declaration – electromagnetic immunity
The 658 is intended for use in the electromagnetic environment specified below.
The customer or the user of the 658 should assure that it is used in such an environment.
Immunity test IEC 60
601 C
ompliance Electromagnetic environment
Test level level guidance
Electrostatic
Discharge(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
±8 kV contact
±2kV, ±4kV,
±8kV, ±15kV air
±2 kV for power
supply lines
±0.5kV, ±1kV
0%UT
For 0,5 cycle
0%UT
1 cycle
70% UT
(30% dip in UT)
For 25 cycle
0%UT
For 5 c
ycle
3
0A/m
±8 kV contact
±2kV, ±4kV,
±8kV, ±15kV air
±2 kV for power
supply lines
±0.5kV, ±1kV
0%UT
For 0,5 cycle
0%UT
1 cycle
70% UT
(30% dip in UT)
For 25 cycle
0%UT
For 5 cycle
30A/m
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Mains power quality should be that of a typical
comme
rcial or hospital environment
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the user of
658 requires continued operation during power
mains interruptions, it is recommended that the
658 be powered from an uninterruptible power
supply or a battery.
Power frequency magnetic fields should be at lev-
els characteristic of a typical location in a typical
commercial or hospital environment
Guidance and manufacturer’s declaration – electromagnetic immunity for MEDICAL ELECTRICAL
EQUIPMENT and MEDICAL ELECTRICAL SYSTEMS that are not LIFE-SUPPORTING.