Product Manual

Periodically check that eartips are tightly secured to binaurals.
Do NOT immerse stethoscope in any liquid or subject to steam sterilization.
Chestpieces and diaphragms can be wiped down with 70% isopropyl alcohol for
cleaning purposes.
It is unnecessary to disassemble the unit for cleaning. Use a slightly moistened cotton cloth with al-
cohol to clean chestpiece diaphragm, binaurals, and eartips. Be careful to prevent alcohol seepage
u
nder the diaphragm. It may cause significant damage to the microphone and electronic circuits.
Caution - Never immerse instrument in any liquid. Do not apply excessive force to the di-
aphragm. Your Adscope
TM
658 electronic stethoscope has a replaceable diaphragm with a
piezopolymer film contact microphone positioned behind it. The chestpiece is not user-serviceable
and should never be disassembled, except to replace the diaphragm.
To order any replacement parts, contact our Customer Service department at 1-800-232-2670.
Troubleshooting Possible Problems
Problem Solution
Battery will not charge
Poor sound quality, intermittent
sound amplification, or
no sound amplification
Damaged/worn chestpiece
diaphragm
Low battery icon is displaying
Check battery connection. If battery is properly connected and will not
charge, battery must be replaced. Contact your local ADC dealer.
Contact your local ADC dealer for warranty service on your stethoscope.
Contact your local ADC dealer for warranty service on your stethoscope.
Charge battery. If icon displays frequently despite regular charges contact
your local ADC dealer for replacement battery.
Troubleshooting Solutions
ITEM CONTENT
Sensor: Piezoelectric sound transducer activated by polycarbonate diaphragm
Voltage: 3.7v Li-polymer
Power Dissipation: 0.115VA (3.7v * 31mA)
Microphone 5Hz - 5,000Hz
Filter Range: Bell Mode 15Hz - 200Hz
Diaphragm Mode 100Hz - 500Hz
Battery Life 20 hours under continuous power on
Charging Time 70 minutes
Amplifier Amplification range 6dB - 18dB
Control Volume
C
ontrol Volume Digital volume (8 step), 3dB/step
Filter Range
Specifications