Operation of the Adash 3600-MAIN Module
If the unit is in operation, it displays continuously measurement results and, based on the results,
updates the output current of current loops 4 – 20 mA in accordance with their set measurement
range. At the same time, status indicators M (measurement), S (measurement parameters setting) and
C (communication between the modules of the set) flash.
Also the measured signal values are checked continuously for the set alarm limits overload and, based
on the evaluation, the relative relays switch. If no alarm limit is exceeded, the OK relay is active. If any
alarm limit is exceeded, the OK relay opens and the relay that signals the state of the relative
exceeded limit becomes active.
If no error is detected, the SYSTEM relay closes. If communication failure, sensor supply failure etc.
occur, indicator E switches on and the type of error appears on the display. Simultaneously, the
SYSTEM relay opens.
Errors Displayed by the Adash 3600-MAIN Module
ERR ICP - sensor supply error, sensor disconnected or short-circuited, measurement
impossible.
ERR OVR - sensor overexcited, vibration value exceeded the measurement range,
measurement impossible.
ERR NET - data buffer memory for network communication is full, current values cannot be
stored.
ERR MEM - no response from the Adash 3600-MEM module, usually the flash card is removed.
ERR STO - data storage error, usually the STOP delay of the Adash 3600-MEM module is
counting down.
ERR xx - error of communication with module No. xx (each module has its No. xx on the
system interface)
List of Module Numbers on Interface RS-485
In case of error, the module number appears on the Adash 3600-MAIN display in the HEX form.
Module number Module type
--------------------------------------------------------------------------------------
11H to 1BH Adash 3600-REL, module 1 to 12 respectively
21H to 2BH Adash 3600-MPX, module 1 to 12 respectively
31H to 3FH Adash 3600-LOOPS, module 1 to 15 respectively
51H Adash 3600-COM, COMD or NET
53H Adash 3600-GSM
55H Adash 3600-MAIN
57H Adash 3600-MEM