Measurement by an External Analyser
1. To connect an external analyser to the Adash 3600 set, the Adash 3600-COM module is equipped
with a BNC connector marked SIGNAL, to which the measured signal is fed from the vibration sensor.
This signal is fed to the connector from the AC OUT and AC GND terminal connectors of the Adash
3600-MAIN module and is already free of the DC component of ICP supply.
Fig. Measurement terminal
2. If you connect the analyser directly to the +ICP and -ICP terminal connectors of the Adash 3600-
MAIN module, you measure the signal from the vibration sensor directly without any conditioning,
therefore including the DC component of ICP supply (approx. 12 V).
System Extension
The basic single channel Adash 3600 set can be extended to a multi-channel set by connecting the
Adash 3600-MPX modules. The multi-channel sets can be further completed with other modules,
which will assure the display, archiving and processing of the measured data.
Multi-channel Set
A multi-channel set can be created from the basic Adash 3600 set by extending it with 1 to 8 Adash
3600-MPX modules. Up to 8 vibration sensors can be connected to each multiplexer module, up to
four measurement types can be requested at each sensor via 4 signal paths LF, HF, LIN and ENV.
The configuration of the requested measurement types at individual vibration sensors is performed via
the Adash 3600 Setup program.
Network On-line Data Transmission
If the Adash 3600 set is mounted with the Adash 3600-NET network communication module (it
replaces the Adash 3600-COM module in the set), on-line transmission of the measured data to a host
PC via LAN network is enabled.
For archiving and processing of the measured data the DDS 2000 user software by Adash is required
on the host PC. If also there is on-line measured data to be processed, the user software must be
complemented with DDS Data Manager, which controls the data transfer from the A3600 to the PC.
NOTE: If you are NOT using a multiplexer unit
Before connecting the analyser to the SIGNAL connector of the
Adash 3600-COM module, put a measurement terminal on its
second connector marked SERIAL, which will assure the
interruption of measurement at the Adash 3600-MAIN module
and the outlet of the measured signal with amplification 1
to connector SIGNAL.