Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 88
Servicing Instructions 227 436 37 Rev. B V1.3
Electrical Diagram for Patient Leakage Current Test
For protection of the test person, the following values of resistor R may be used:
Typ BF 22 kOhm (120 to 130 V)
47 kOhm (220 to 240 V)
Typ CF 100 kOhm (220 to 240 V)
9.3 Maintenance, cleaning, disinfection
CardioSys maintenance, cleaning, disinfection is performed in accordance with the
User’s Manual.