Operating instructions
Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 78
Servicing Instructions 227 436 37 Rev. B V1.3
[17627] Zero adjustment not possible
PRESS only; signal is too unstable
[17628] Sensor unconnected
[17629] Interference in signal/course of the curve
[17630] Trigger point cannot be found
for HMV only; injection must be made within 20 s and the duration of actual
injection must not exceed 4 s
[17631] Basal blood temperature fluctuates
[17632] Incorrect injectate temperature
[17633] Signal modulation too small
for blood pressure and pulse rate;
[17655] Incorrect blood temperature
8.5 General Troubleshooting
- Stress test driver settings lost
Stress test profiles were changed or newly created and are no longer there.
Remedy: The file PERM_MEM.DAT in the Windows directory was inadvertently deleted
or overwritten. A backup of this file should be made after changing profiles so that it can
be restored even after a hard disk error (restore file into Windows directory).
- Defective data sets in the database
Errors occur during patient selection, but not when selecting examination options.
Remedy: Contact technical support for a reorganisation tool, which may repair the error
in the data base.
By activating the REORG procedure you can reorganize defective BTRIEVE files from
CardioSoft. This repair procedure may become necessary when the data become
damaged for some reason or another (e.g. after a power failure). To carry out the repair
procedure successfully, CardioSys must first of all be terminated. There must be
sufficent memory space on your hard disk.
As a general rule, the following applies:
memory space required = twice as much free memory space as the sum of all
The original CardioSys files are still available in the form of “OLD” files after
reorganization and should only be deleted when the reorganization has been completed
successfully.
If the reorganization is unsuccessful, the only option remaining is to retrieve the latest
backup data.
- There are problems with my printer, since I connnected CORINA to the system