Operating instructions

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 76
Servicing Instructions 227 436 37 Rev. B V1.3
[13236] No examination data are stored on recorder!
Examination data retrieval only possible when a recording was made with
TONOPORT.
[13237] Number of retrieved data defective!
Data transmission from TONOPORT to CardioSys was defective.
Remedy:
- Retrieve examination data a second time
[13238] Error when retrieving the file "LBD_TEST.PAT" in the program directory!
Test patient data are stored in the file "LBD_TEST.PAT" in the program
directory. The file is either damaged or does not exist.
Remedy:
- Check hard disk.
[13239] Error when saving examination!
A hard disk error has occurred when saving the examination.
Remedy:
- Check hard disk.
[13241] Error when retrieving the examination file!
The examination file is damaged
Remedy:
- Restore examination data backup. However, when doing this all the
examinations performed after this backup get lost !!!
[13243] Programing not possible, recorder still has stored examination data!
The TONOPORT recorder can only be programmed when it does not have
any stored examination data.
Remedy:
- Clear TONOPORT data memory
[13274] Not all data could be downloaded. Patient monitor was too often turned on
and off!
The TONOPORT recorder was more than 15 times switched off and on again.
8.4.11 Error messages [14xxx]: Late potentials
[14055] Recording only possible when you have the option 'AVER: Late Potentials'.
8.4.12 Error messages [15xxxx]: General tools
[15004] Error when opening signal processing
[15005] CORINA error: data transmission
see error message [6556]
[15008] Message: "Saving" is only possible when you have the option "Network"!
[15009] Insufficient free memory space on hard disk
[15011] Internal CORINA error
see error message [6509]