Operating instructions

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 75
Servicing Instructions 227 436 37 Rev. B V1.3
Remedy:
- Restore examination data backup. However, when doing this all the
examinations performed after this backup get lost !!!
[8545] Read examination file : Error when retrieving the file
The content of the examination file has probably been damaged.
Remedy:
- Restore examination data backup. However, when doing this all the
examinations performed after this backup are lost !!!
[8551] Respiratory flow sensor not connected to selected interface or switched off
Remedy:
- Check whether the respiratory flow sensor is switched on. If it is switched on,
check whether the respiratory flow sensor is connected to the correct
interface. If both are OK, the respiratory flow sensor or the interface could be
defective.
8.4.7 Error messages [10xxx]: Ergometry monitoring
[10427] No (further) ergometry for monitoring available!
[10428] Ergometry test terminated!
8.4.8 Error messages [11xxx]: Printer tools
[11007] Error when initializing the print job
Possibly too little space free in main memory.
Remedy:
- Exit Windows and restart.
[11008] Error when starting the print job
Too little space in main memory to transmit the print data (spooling).
Remedy:
- Exit Windows and restart.
8.4.9 Error messages [12xxxx]: Printer tools
[12064] Local printer is not ready!
[12065] No ST measurement at this time!
[12066] Error when starting the print job!
[12500] Error when saving the episode
8.4.10 Error messages [13xxx]: Long-term blood pressure
[13130] Error when initializing the interface!
Occurs when the configured interface is already busy, for example.
Remedy:
- Select another free COM interface.
[13228] Error during data transmission from/to TONOPORT
Remedy:
- Check TONOPORT - CardioSys connection. Is TONOPORT switched on
and is battery still sufficiently charged?