Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 74
Servicing Instructions 227 436 37 Rev. B V1.3
- Check option code number under General Settings, options for
ergometry and correct if necessary.
[07984] Storage only possible when you have the network option!
The examination can only be saved on the network server when the network
software option has been purchased.
Remedy:
- Operate program locally
- Purchase network software option
[07993] Error in CORINA data transmission!
see error message [6556]
[07994] Error in CORINA downloading!
see error message [6555]
[07995] Internal CORINA error!
see error message [6509]
[07996] Error when opening network
An error has occurred when logging into the NetBios network.
Remedy:
- This error can occur after entering a valid station name and the network
monitoring system has been activated during ergometry recording. When the
ergometry recording is called up and the program CardioSys/ CardioSoft is
terminated immediately afterwards, the next time the ergometry recording
mode is called up this leads to this error. If the system is switched off after
quitting the program and switched on again, this error does not occur. This
error can also be 'repaired' by entering a new station name in the interim. The
error only occurs when less than approx. 10 have passed between calling up
the ergometry recording and terminating the program. Cannot be corrected by
technical means!
8.4.6 Error messages [8xxx]: Spirometry
[8525] The examination has not been assigned to a patient; thus, it is not possible to
enter an interpretation!
[8543] Read examination file: Error when opening file
The examination file does not exist or the user does not have any read rights
(can only occur in network).
The rights of the user in the network should be checked. If the user has read
rights for the subdirectories with the examination data (as a rule he should have
read and write rights for the CardioSoft installation directory, including all the
subdirectories) or CardioSoft is installed locally, the examination file has been
deleted or the data base damaged.
Remedy:
- Restore examination data backup. However, when doing this all the
examinations performed after this backup get lost !!!
- Under NetWare flag is not set on S (Sharable). See NetWare
Documentation.
[8544] Read examination file: No spirometry file
The data base has probably been damaged.