Operating instructions

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 73
Servicing Instructions 227 436 37 Rev. B V1.3
There are permanent and adjustable load profiles in the ergometer.
Permanent load profiles cannot be modified or deleted.
Remedy:
- Load a permanent load profile and press the “new” key. This copies
the permanent profile and can now be modified and saved under a
new name.
[07657] Load profile cannot be modified!
[07961] NetBios driver not loaded!
The NetBios driver must be loaded in the ergometry recording station and the
slave screen station to enable the ergometry slave screen function.
Remedy:
- Load NetBios driver before starting Windows and
CardioSys/CardioSoft
[07962] No station name entered in General Settings!
To enable the ergometry slave screen function an unambiguous station name
for all ergometry recording stations and ergometry slave screen stations must
be entered under General Settings in the network.
Remedy:
- Enter unambiguous station name under General Settings in the
network
[07963] No connection to ergometer / blood pressure meter!
There is no connection between the ergometer and the PC.
Remedy:
- Has the correct connection cable been used
- Is the cable connected up correctly
- Is the ergometer switched on
- Is the ergometer a Marquette Hellige ergometer (ID)
[07964] CORINA not connected or not switched on!
see error message [6100]
[07965] Insufficient free memory on hard disk!
The hard disk does not have enough free memory space. CardioSys already
includes a certain number of bytes in hand.
Remedy:
- Make memory space available on the hard disk, if necessary, store
examinations externally
- If online ECG has been configured, the memory requirement can be
reduced by disabling this option.
[07966] Error when saving the examination!
An error has occurred when saving the examination on the hard disk.
[07967] Error when saving the online ECG!
An error has occurred when online saving ECG data on the hard disk.
[07983] No ergometry software option available!
No software option was procured for the selected action. Ergometry cannot be
called up.
Remedy:
- Purchase Ergometry software option