Operating instructions

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 72
Servicing Instructions 227 436 37 Rev. B V1.3
[6508] Error when opening signal processing
[6509] Internal CORINA error
CORINA has detected an internal error during the self-test.
Remedy:
- Replace CORINA
[6510] Error when retrieving the examination data
The content of the examination file has probably been damaged.
Remedy:
- Restore examination data backup. However, when doing this all the
examinations made after this backup get lost !!!
[6511] Error when retrieving control data
see error message [6510]
[6515] Message: "Interpretation" is only possible when you have the option
"Measurement"!
[6517] There is too little space on your hard disk to save this recording
[6518] Error whilst saving the online ECG
When saving the data an error occurred on the hard disk. (Disk full or damaged)
[6539] Message: "Save" is only possible when you have the option "Network"!
[6540] Error when writing the examination data
The hard disk is probably damaged or full. The examination must be repeated.
[6545] Function disabled: another user is changing the interpretation
[6546] Function disabled: another user is performing a reanalysis
[6547] Function disabled: another user is viewing this examination
[6548] Function disabled: another user is entering a comment on this examination
[6549] Message: The channel designation for both recordings differs. A comparison of
corresponding channels is thus not possible!
[6555] Error during CORINA downloading
An error occurred when updating internal CORINA software.
Remedy:
- Restart program
[6556] CORINA error: data transmission
Errors occurred during data transmission from CORINA to the PC
Remedy:
- Check connection from CORINA to PC
8.4.5 Error messages [7xxx]: Ergometry
[07327] Error when retrieving the examination
Examination file damaged.
Remedy:
- Restore examination data backup. However, when doing this all the
examinations made after this backup are lost !!!
[07436] Max. number of characters per line exceeded
[07438] Max. number of lines exceeded
[07654] Number of load parameters too high!
[07655] No further load stage possible!
[07656] Load profile cannot be modified!