Operating instructions

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 71
Servicing Instructions 227 436 37 Rev. B V1.3
[2158] No examinations found on data carrier
[2161] Another network user is making a sound recording
[2166] The selected examination does not exist on the data carrier
[2168] This function cannot be executed in the DEMO version
[2169] Error whilst writing on the data carrier for external storage
[2170] Error whilst reading from the data carrier for swapping in
[2321] Maximum number of sound recordings attained
8.4.3 Error messages [3xxx]: Main program
[3421] Path or file name not found!
[3422] Error when executing the file:
[3423] Insufficient free memory! Please terminate the program and Windows and then
restart both. Further program execute commands may lead to errors!
[3425] Insufficient free resources. Please terminate the program and Windows and then
restart both. Further program execute commands may lead to errors!
[3426] When operating an external recording mode a patient must be selected!
[3428] No external operating mode configured for this examination!
[3432] Error during transfer of DDE command!
[3433] No examination data from external program!
8.4.4 Error messages [6xxx]: Resting ECG, Emergency
[6060] Function inoperative as examination was not assigned to a patient.
[6062] Message: Storage time [min:sec]
During an emergency recording, this displays how many minutes and seconds
are still available on the hard disk to save the online ECG.
[6100] CORINA not connected or not switched on!
The programm cannot establish any contact to CORINA.
Remedy:
- Connect up CORINA power supply unit
- Connect CORINA to an LPT on PC
- Switch CORINA briefly off then on again (disconnect power unit) and try again
- in BIOS change the parallel interface mode from 'Bidirectional' to 'AT or
unidirectional'
- CORINA is inoperative with the parallel interface of IBM PSx PCs; these PCs
require the additional installation of a normal parallel interface card
[6271] No data available
[6400] Message: "Recording” is only possible when you have the option "Resting ECG
Basic Features"!
[6467] At least one of the options "10-s ECG Analysis" or "Online Storage" must have
been selected!
[6480] Message: "Recording” is only possible when you have the option "Resting ECG
Basic Features"! or "Online ECG Storage"!
[6494] The maximum limit for %d lines has been reached!