Operating instructions

Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 70
Servicing Instructions 227 436 37 Rev. B V1.3
[1156] Error when decompressing data
The selected examination file has been damaged.
Remedy:
- Restore examination data backup. However, when doing this all the
examinations performed after this backup get lost !!!
[1161] This function cannot be executed in combination with a physician accounts
program
[1162] This function cannot be executed in the DEMO version
The limitations of the DEMO version are described in the UM.
Remedy:
- Use complete version of the program.
[1163] No patient selected for the examination by data carrier
[1174] No Patient selected for the examination by ECG recorder
[1176] Patient identification not clear. Patient data cannot be used.
[1178] Software option request list unavailable
[1180] No request for selected examination mode available
[1181] Warning: Only the latest 20 examinations are displayed
[1254] No examination selected
[1517] No modem interface or modem busy
Mouse interface was selected.
Remedy. Select another interface.
[1531] Defective initialization command for modem.
Wrong modem model selected.
[1534] Distant end busy
Wrong dialing procedure programmed and telephone system gives engaged
[1544] Error during data transmission
Bad telephone connection or no FIFO chip (16550) on serial interface
Remedy: Dial again; install FIFO chip.
8.4.2 Error messages [2xxx]: Help functions
[2129] Wrong password entered
[2135] Settings cannot be loaded
The INI file in the directory \cardio\setup has been damaged.
Remedy:
- Overwrite settings in this entry with new settings using the help function
“Save Settings” (Warning: damaged settings are lost)
[2136] Setting cannot be saved
The INI file in the directory \cardio\setup has been write-protected or the hard
disk is full
Remedy:
- Disable INI file write protection in the directory \cardio\setup using the file
manager or create memory space on the hard disk.
[2144] Invalid date
[2145] No examination fulfils compressing criterion
[2147] No examination fulfils external storage criterion
[2157] The examination on the data carrier is not from this program