Operating instructions
Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 69
Servicing Instructions 227 436 37 Rev. B V1.3
[1115] General data base error (No.%s)
An error occurred when entering a new patient into the data base file in the
directory \cardio\database.
Remedy:
- Restore examination data backup. However, when doing this all the
examinations performed after this backup get lost !!!
[1116] Date of birth should be entered in prescribed format
[1132] Examination cannot be compared with itself!
[1134] Error when accessing data base files
The data base files in directory \cardio\database have been damaged or deleted.
Remedy:
- Restore examination data backup. However, when doing this all the
examinations performed after this backup get lost !!!
[1135] Error when accessing BDT file (%s)
The BDT file from the physician accounts program is unavailable or the directory
given for this BDT file is incorrect.
Remedy: Retrieval in physician accounts program improved.
[1136] Error when reading from the BDT file in line %s
There is an error in the line given of the BDT file from the physician accounts
program (e.g., field length is incorrect)
Remedy:
- The physician accounts program must create BDT file correctly.
[1138] Connection busy or not available
[1142] No patient has been selected for the transfer of examination data
[1145] Error when copying the examination
An error occurred when copying the examination from/onto storage medium. If
copying onto a data carrier, then the data carrier was defective. If copying from a
data carrier, the hard disk is full.
Remedy:
- Using a non-defective data carrier or create free memory on the hard disk.
[1149] Only capital letters (A to Z), numbers (0 to 9) and special characters
($,/,.,+,-,%,space) can be used in patient IDs
[1150] Another user is accessing this patient’s data in the network
Another user is performing this examination.
Remedy:
- Terminating the program at the other network stations.
- If this error message appears even when no other network station is busy, exit
program and delete file “network.btr” in the installation directory (e.g.,
Y:\CARDIO) in subdirectory "DATABASE".
[1151] This examination is being performed by another network user
[1152] Another user is accessing one of the patient’s examinations in the network
[1153]Selected patient cannot be deleted.
[1154] Selected examination does not exist
[1155] Error when compressing data
The hard disk is full.
Remedy:
- Create memory space on the hard disk, e.g., by storing examinations
externally.