Operating instructions
Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 66
Servicing Instructions 227 436 37 Rev. B V1.3
For Windows 95/98:
With the Registry Editor please locate the following key:
"HKEY_LOCAL_MACHINE\SYSTEM\CurrentControlSet\Services\Class\netTrans\000n"
Where 000n refers to the network adapter that TCP/IP is bound to.
Please enter the key "MaxMTU" with the Type "String" with the
value "512" (Important: use Type "String" and enter the string "512").
After the change please reboot the System.
--------------------------------------------------------------------------------------------------
Question: CORINA is not detected at the LPT port:
Solution:
Set LPT Mode in BIOS to „NORMAL MODE“ (not to „ECP“ or „EPP“ Mode)
If this will not help, please use Multi I/O card from Hellige Service
(PN: 2000148-001)
There will be a new CORINA generation with better communcation capabilities
(CORINA V3.1). Because of the enhanced communication features, this new
CORINA only will work with newer CardioSoft versions greater/equal V4.14
--------------------------------------------------------------------------------------------------
Question: I got the error message -115 when installing CardioSoft V4.1x
Solution:
The file that was displayed within the error message is write protected or the actual User
do not have the rights to write to the destination directory. Change the readonly attribute
for the file or the User’s policies.
Question:
When I start CardioSoft, I get a data base error
Solution:
Copy new „REORG.BAT“ file from CardioSoft V4.1x directory to „DATABASE“ directory
and start a reorganisation of the data base.
--------------------------------------------------------------------------------------------------
Question: When I open an examination I get a Decompression Error:
Solution:
Use Check and repair hard disk using SCANDISK or CHKDSK. Then restore the system
By using the last BACKUP file.
--------------------------------------------------------------------------------------------------
Question: PCW130 value will not be displayed in the stress test measurement
summary
Solution:
PCW 130 is not for use in combination with treadmill stress tests.
The stree test must be done, weight of the patient is existing and a heart rate of 130 was
reached. (ECG equipment must have the capability for a hart rate of 130)